PC3 Peltier Chiller
The PC3 is a compact Peltier Cooled inlet system which incorporates the ESI cyclonic spray chamber. The peltier within the PC3 is air cooled, as such the system can be connected to any ICP-OES or ICP-MS. The PC3 cools the outer walls of the cyclonic spray chamber reducing the water or solvent vapor loading on the plasma resulting in enhanced stability and performance.
- Thermal stabilization of spray chamber improves long-term stability
- Default settings of spray chamber temperature (+2°C/ -5°C) for aqueous or organic solvents. Custom temperature settings are available upon request.
- Fast rinse-out using PFA-ST nebulizer and O-ring-free quartz cyclonic spray chamber.
- No separate chiller required
- No antifreeze required
- No algae growth
- No water lines
- Robust, with very little maintenance reducing the chance of accidental breakage
The PC3-SSI incorporates a dual spray chamber which further homogenizes the sample aerosol stream giving improved stability. Ideal for high precision Isotope Ratio analysis.
Isotope |
|
PC3 |
PC3-SSI |
Co-59 |
1.03% |
0.92% |
0.53% |
In-115 |
1.32% |
1.08% |
0.40% |
Ba-138 |
0.96% |
0.90% |
0.48% |
U-238 |
1.16% |
0.93% |
0.34% |
Short term signal stability, 20 min analysis time n = 10
|
|
 |